6. Commissioning, operation

6. Commissioning, operation

6.1 Electrical connection

The connection of the attachable indicator is made via simply plugging it

between an existing transmitter and connector using a specific adapter

design for connectors to DIN 175301-803 A.

No separate power supply is needed since the instrument is powered directly

from the measuring current.

The connection and commissioning should only be carried out

by skilled, qualified personnel. An incorrect connection could

damage the indicator.

CAUTION!

Note the maximum permissible input current of 40 mA!

6.2 Pin assignment

The assignment of the angular connector is designed for the most common-

ly-used configuration for the respective input signal. Since this assignment

is not standardised, it can happen that the assignment for the connected

transmitter does not match the assignment of the attachable indicator.

Standard assignment for the model A-IAI-1 angular connector

Contact no.

Wiring colour Pin

Female connector

1

grey

1)

display +

display -

2

red

connected

3

black

connected

4

yellow

connected

1) The wire between pin 1 and the device is blue, the wire between the device and female contact 1 is

grey. Only the grey wire can be changed.