GB

WIKA operating instructions attachable indicator model A-IAI-1

18

14016825.01 02/2012 GB/D

Once the cover of the case has been removed, to configure the instrument

continue as follows:

To access the menu, while the actual value is displayed, press key 2 for 2

seconds until "dP" appears in the display.

To set the parameter values, press keys 2 and 3.

To save the set value, press key 1. The parameter name will then appear in

the display once more.

To switch to the next parameter, press key 1. The name of the parameter

will appear in the display.

If, during input, no key is pressed for 60 seconds, the configu-

ration of the instrument will be interrupted. Previously-saved

values will not be lost.

Parameters of the A-IAI-1 attachable indicator

Parameters Values

Description

Key 1

Keys 2 and 3

dP

Position of the decimal point (decimal point)
----

Max. indication range: -1999 … 9999

---.-

Max. indication range: -199.9 … 999.9

--.--

Max. indication range: -19.99 … 99.99

-.---

Max. indication range: -1.999 … 9.999

di.Lo

Lower indication range limit (display low)
-1999 ... 9999 This value will be displayed when the input signal

= 4 mA

di.Hi

Upper indication range limit (display high)
-1999 ... 9999 This value will be displayed when the input signal

= 20 mA

Li

Measuring range limit (limit)
oFF

deactivated

Exceeding the measuring range limits is permitted

up to the measuring limit (see note)

on.Er

active, (error display):

The measuring range is limited exactly to the input

signal. When this exceeds or drops below it, a

corresponding error message is displayed.

6. Commissioning, operation