WIKA operating instructions attachable indicator model A-IAI-1

20

14016825.01 02/2012 GB/D

6. Commissioning, operation / 7. Maintenance and cleaning

Parameters Values

Description

Key 1

Keys 2 and 3

ScAL

Slope (scale)
-5.00 … 5.00 The input of the slope correction factor is made

in %. The displayed value is calculated as per the

following equation:

Display = (measured value - Offset - di.Lo) * (1 +

slope correction [% / 100] ) + di.Lo

Example for offset and slope correction

Connection of a pressure transmitter

The instrument display without offset and slope correction is as follows:

at 0 bar = 0.08, at 20 bar = 20.02

From this is calculated:

Zero point:

0.08

Slope:

20.02 – 0.08 = 19.94

Deviation:

0.06 = set slope - actual slope = 20.00 – 19.94

Therefore the following should be set:

Offset =

0.08 = zero error

Scale =

0.30 = deviation / actual slope = 0.06 /19.94 = 0.0030 = 0.30 %

Min/Max value memory

The instrument has a Min/Max value memory. In this are stored the lowest

and the highest diplayed values. The Min/Max value memory is a volatile

memory, meaning the stored values are lost when the power is interrupted.

Key

Display

To recall Min

value:

Press key 3 briefly

"Lo" will be displayed briefly and then for

approx. 2 s the Min value will be shown.

To recall Max

value:

Press key 2 briefly

"Hi" will be displayed briefly and then for

approx. 2 s the Max value will be shown.

To delete

the Min/Max

values:

Press keys 2 and

3 simultaneously

for 2 s.

"CLr" will briefly be shown in the display,

the Min/Max value will be reset to the

current display value.