• Pull forward and turn the locking
button to adjust the working
pressure until the manometer
indicates 3 bar. During this
procedure press down the release
lever (6) at the right side of the
instrument several times in order
to guarantee a precise pressure
adjustment.
• Push back the button (1) to re-
lock the pressure reducer. Now
the unit is ready for operation.
3. Adjusting the Working
Pressure of 3 bar
• Connect the open end of the air
hose to the fitting piece on the
back panel of the instrument
(R
1
/
8
") and secure the air hose
with the union nut.
• Connect the other end of the
hose, equipped with a fitting
(NW 7.2 mm), to the quick action
closure of the compressed-air
supply (max. 5 bar).
• The compressed air should be
oilfree and dehumidified by a
water seperator.
• If the air supply has to be stopped
or interrupted, first release the
quick action closure of the
compressed-air supply (which
should be near the instrument),
and then release the union nut.
2. Connection to the
Compressed-air System