9. Maintenance and
Cleaning
Remaining paint chips at the
opening need to be removed
frequently with a wet cloth.
Do not use solvents.
Error description:
• Impact does not appear
homogenious:
Check impact tool for damage
and remaining paint scrap. Please
refer also to chapter: "Exchange
of impact tool".
• Impact does not penetrate
through substrate or is not clearly
perceirable: Check air pressure
gage (3 bar).
• Pressure loss in account of a
leakage inside of the instrument:
Please contact BYK-Gardner
Technical Service.
10. Components
esp-10 Single Impact Tester
Weight
Connection hose
Operating instructions
10. Ordering Guide
Cat.No.
Description
5200
esp-10
Single-Impact Tester
5201
Impact Tool,
Replacement
5202
Weight,
Replacement
5824
Precision Microscope
5205
Calibration Standard
Disposal:
• The esp-10 and all of it's parts are
to be disposed according to
material or labeling in conside-
ration of legal requirements.