WIKA operating instructions temperature transmitter, model T19
3. Configuration
3.1 Working sequence
Model T19.10 head mounting:
Remove case bottom
Model T19.30 rail mounting:
Remove case lid
Set the solder bridges for the
desired measuring range in
accordance with tables 1 – 7. The
measuring ranges to be selected
depend on the model of the
transmitter.
Model T19.10: Snapfit the bottom to the case again
Model T19.30: Snapfit the lid to the case again. Insert lid with its long
side into the case, then press lid to the case until it locks into place
Adjust zero point and span by means of potentiometer as described
in section 7!
Note the measuring range on the rating plate (for example, with
waterproof fibretip pen)
3. Configuration
Demonstration: Model T19.30
use a small screwdriver:
to prick the marking
unlook both locks of the snap-in
connection