WIKA operating instructions model TR30

11377551.06 07/2014 GB/D

13

Electromagnetic compatibility

(EMC)

2004/108/EC, EN 61326 emission (Group 1,

Class B) and interference immunity (industrial

application)

4)

Electrical connection

M12 x 1 circular connector, 4-pin

Angular DIN connector Form A for cable with

Ø 6 ... 8 mm, cross section max. 1.5 mm²

Readings in % refer to the measuring span

For a correct determination of the overall measuring error, both sensor and transmit-

ter measuring deviations have to be considered.

1) For detailed specifications for Pt100 sensors, see Technical Information IN 00.17 at www.wika.com.

2) The temperature transmitter should therefore be protected from temperatures over 85 °C

3) For measuring spans smaller than 50 K additional 0.1 K

4) Use resistance thermometers with shielded cable, and ground the shield on at least one end of

the lead, if the cables are longer than 30 m or leave the building.

CE conformity, approvals

CE conformity

EMC directive

5)

2004/108/EC, EN 61326 emission (group 1, class B) and

interference immunity (industrial application)

ATEX directive

94/9/EG (Option)

Approvals (option)

NEPSI

Ignition protection type “i” - intrinsic safety,

ignition protection type “iD” - dust protection through

intrinsic safety, China

EAC

Import certificate, ignition protection type “i” - intrinsic

safety, ignition protection type “iD” - dust protection

through intrinsic safety, customs union Russia/Belarus/

Kazakhstan

GOST

Metrology/measurement technology, Russia

DNV

Ships, shipbuilding (e.g. offshore), International

5) Only for built-in transmitter

For further specifications see WIKA data sheet TE 60.30.

3. Specifications