The lower limit of the measuring range is configurable between

-50 ...+150 °C.

The possible upper limit of the measuring range is dependent on

the respective lower limit of the measuring range. For possible

combinations of lower and upper limits of the measuring range, see

diagram. For clarity, this dependency has been shown in 50 °C steps

in the examples in this diagram. The configuration software checks

the required measuring range and will only accept permissible values.

Intermediate values are configurable; the smallest increment is 0.1 °C.

The thermometers are delivered with a basic configuration (0 ... +150 °C,

downscale) or configured to customer specifications within the

configurable limits.

With configuration to customer specifications, the measuring range

will be shown clearly on the instrument label. The configuration and

the choice of parameters is described in the menu-driven configuration

software. Changes to the configuration must be noted on the label using

a water-resistant felt-tip pen.

Possible combinations of the lower and upper limits of the

measuring range

The upper limit of the measuring range is dependent on the respective

lower limit of the measuring range. For clarity, this dependency has

been shown in 50 °C steps in the examples in these diagrams. The

configuration software checks the required measuring range and will

only accept permissible values. Intermediate values are configurable; the

smallest increment is 0.1 °C.

7. Configuration model TR30-W