Gb 3. specifications – WIKA CEP3000 Benutzerhandbuch

Seite 11

Advertising
background image

11

WIKA operating instructions, model CEP3000

14036068.02 08/2012 GB/D

GB

3. Specifications

Input and output

signals

Measuring

range

Absolute measurement

uncertainty (all errors

incl.)

Resistance thermometer

1)

Pt100 (385)

-200.0 ... +800.0 °C

0.33 °C

Pt100 (3926)

-200.0 ... +630.0 °C

0.3 °C

Pt100 (3916)

-200.0 ... +630.0 °C

0.3 °C

Pt200 (385)

-200.0 ... +630.0 °C

0.8 °C

Pt500 (385)

-200.0 ... +500.0 °C
500.0 ... 630.0 °C

0.3 °C
0.4 °C

Pt1000 (385)

-200.0 ... +100.0 °C
100.0 ... 680.0 °C

0.2 °C
0.3 °C

Ni120 (672)

-80.0 ... +260.0 °C

0.2 °C

Cu 10

-100.0 ... +250.0 °C

2.2 °C

Cu 50

-180.0 ... +200.0 °C

0.5 °C

Cu 100

-180.0 ... +200.0 °C

0.3 °C

YSI 400

15.0 ... 50.0 °C

0.1 °C

Resistance

Excitation current

Output

5.0 ... 400.0 Ω
5.0 ... 400.0 Ω
401 ... 1,500 Ω
1,501 ... 3,200 Ω

0.15 Ω
0.1 Ω
0.5 Ω
1.0 Ω

0.1 ... 0.5 mA
0.5 ... 3 mA
0.05 ... 0.8 mA
0.05 ... 0.4 mA

Input

0.00 ... 400.00 Ω
401.0 ... 1,500.0 Ω
1,501.0 ... 3,200.0 Ω

0.1 Ω
0.5 Ω
1.0 Ω

1) Absolute measurement uncertainty based on a 4-wire circuit

For further specifications see WIKA data sheet CT 82.01 and the order documentation.