WIKA operating instructions model CPB3800

14062232.03 12/2013 GB/D

GB

3. Specifications

3. Specifications

Piston-cylinder systems

Measuring range

1)

bar

1 … 120

2.5 ... 300 5 ... 700

10 ... 1,200

Required masses

kg

41

50

58

50

Smallest step

2)

(Standard mass set)

bar

1

2.5

5

10

Nominal cross-sectional area of

the piston

in²

1/16

1/40

1/80

1/160

Measuring range

1)

lb/in² 10 … 1,600 25 ... 4,000 50 ... 10,000 100 ... 16,000

Required masses

kg

47

47

58

47

Smallest step

2)

(Standard mass set)

lb/in² 10

25

50

100

Nominal cross-sectional area of

the piston

in²

1/16

1/40

1/80

1/160

Accuracies
Standard

3) 4)

0.05 % of reading

Option

3) 4)

0.025 % of reading

Pressure transmission

medium

Hydraulic fluid based on VG22 mineral oil (0.5 l

included in scope of delivery)

Material
Piston

Tungsten carbide

Cylinder

Tungsten carbide

Mass set

Stainless steel, non-magnetic

Weight
Piston-cylinder system

kg

2.4

BAR mass set incl. overhang

(bell jar)

kg

41.5

50.5

58.5

50.5

lb/in² mass set incl. overhang

(bell jar)

kg

47.5

47.5

58.5

47.5

Carrying case for mass set

(optional, 2 pieces required)

kg

5.8

Dimensions
Carrying case for mass set

(optional)

400 x 310 x 310 mm (W x H x D) and

215 x 310 x 310 mm (W x H x D)

1) Theoretical starting value; corresponds to the pressure value generated by the piston or the piston and its make-up weights (by

their own weight). To optimise the operating characteristics more weights should be loaded.

2) The smallest pressure change value that can be achieved based on the standard weight set. To reduce this, a set of trim

masses is also available.

3) The accuracy from 10 % of the measuring range is based on the measured value. In the lower range, a fixed error based on

10 % of the range applies.

4) Measurement uncertainty assuming reference conditions (ambient temperature 20 °C, air pressure 1,013 mbar, relative

humidity 40 %). For operation without a CalibratorUnit, corrections must be made if required.