Gb 2. safety 2. safety – WIKA CPB3800 Benutzerhandbuch

Seite 6

Advertising
background image

6

WIKA operating instructions model CPB3800

14062232.03 12/2013 GB/D

GB

2. Safety

2. Safety

WARNING!

Before installation, commissioning and operation, ensure that the

appropriate pressure balance has been selected in terms of measuring

range, design and specific measuring conditions.

Non-observance can result in serious injury and/or damage to the

equipment.

Further important safety instructions can be found in the individual

chapters of these operating instructions.

2.1 Intended use

Pressure balances are the most accurate instruments available on the market for the

calibration of electronic or mechanical pressure measuring instruments. By direct

measurement of the pressure as the quotient of force and area (p = F/A), pressure

balances are approved as primary standards.

The core component of the CPB3800 is therefore a very precisely-manufactured piston-

cylinder system, which is loaded with masses in order to generate the individual test

points. The masses applied are proportional to the target pressure and this is achieved

through graduated masses. A maximum pressure of 1,200 bar must not be exceeded.

The pressure is set via an integrated, finely-adjustable, precision dual area spindle

pump. As soon as the measuring system reaches equilibrium, there is a balance of

forces between the pressure and the mass load applied.Then the test item can be

calibrated or adjustments can be carried out.

Due to its stand-alone operation (integrated pressure generation and the pure

mechanical measuring principle), the model CPB3800 is ideal for on-site use for

maintenance and service.

The instrument has been designed and built solely for the intended use described here,

and may only be used accordingly.

The technical specifications contained in these operating instructions must be

observed. Improper handling or operation of the instrument outside of its technical

specifications requires the instrument to be taken out of service immediately and

inspected by an authorised DH-Budenberg/WIKA service engineer.

Advertising