8. Faults

8. Faults

Faults

Causes

Measures

Equipment does not

provide any output

pressure.

No liquid in pressure balance.

Check that pressure balance is

filled with liquid. Fill the equipment

with fluid as necessary. See chap-

ter 6.5.1 "Filling the equipment

with liquid".

Valve B is open.

Close valve B and try again.

Component being tested has a

large volume.

Pre-fill component with liquid

before test.

Missing or damaged liquid seals

shown by signs of unexplained

liquid leaks.

Examine seals on equipment to

ensure they are fitted correctly

and are undamaged. Replace as

necessary.

Valve B handwheel disconnected

from spindle.

Examine valve B. Tighten up nut

securing handwheel to spindle as

necessary.

Valve B assembly or valve seat

damaged.

Examine condition of valve B and

valve seat. Replace valve assem-

bly or return pressure balance to

DH-Budenberg/WIKA for overhaul

as necessary.

If unable to locate a cause.

Return pressure balance to DH-

Budenberg/WIKA for investigation.

Equipment provides

pressure but pressure

decays to zero

Incorrect operating procedure

being used.

Ensure that correct operating

procedure is being followed (see

chapter 6.6).

Missing or damaged liquid seals

shown by signs of unexplained

liquid leaks.

Examine seals on equipment to

ensure they are fitted correctly

and are undamaged. Replace as

necessary.

Valve A or Valve B valve assembly

or valve seat damaged.

Examine condition of valves A and

B and valve seat. Replace valve

assembly or return pressure bal-

ance to DH-Budenberg/WIKA for

overhaul as necessary.

If unable to locate a cause.

Return pressure balance to DH-

Budenberg/WIKA for investigation.