WIKA operating instructions model CPB3800
14062232.03 12/2013 GB/D
GB
5. Transport, packaging and storage
5. Transport, packaging and storage
5.1 Transport
Check model CPB3800 pressure balance for any damage that may have been caused
by transport. Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair or recalibration).
Masses are shipped in cardboard and not in their respective wooden
cases, if ordered.
Wooden cases are not suitable for use as shipping cases.
5.3 Storage
Permissible conditions at the place of storage:
■
Storage temperature: -10 ... +50 °C
■
Humidity: 35 ... 85 % relative humidity for instrument base and mass set
35 ... 65 % relative humidity for piston-cylinder unit (no condensation)
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Soot, vapour, dust and corrosive gases
■
Potentially explosive environments, flammable atmospheres
■
Corrosive liquids
Store the model CPB3800 pressure balance in its original packaging in a location that
fulfils the conditions listed above. If the original packaging is not available, pack and
store the instrument as described below:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag, containing a
desiccant, inside the packaging.