Comparison Test Pump

GB

CPP 3000-X / CPP 5000-X / CPP 7000-X


WIKA Operating Instructions Comparison Test Pump Version 2.1

16

5.2.2 Filling in of hydraulic oil

.


 Open the high-pressure shut-off valve

 Open the low-pressure shut-off valve

 Turn in the spindle pump clockwise until it reaches the initial stop

 Open the locking screw with the oil filling symbol on top of the basement

 Fill in special oil (0.5 litre supplied, or available as accessory) via the tank opening, until the fill

level reaches the thread of the tank opening (approximately 250ml). The fill level must always be
observed.


 Twist out the spindle pump counter-clockwise until it reaches the rear stop. The filling medium is

suctioned out of the tank into the system.


 Close the tank opening with the locking screw

5.2.3 Venting of the system (after complete filling only)

.


After initial filling, or after a complete oil change, air may be trapped in the system. The system should
be vented using the following procedure:

 Open the high-pressure shut-off valve

 Open the low-pressure shut-off valve

 The test connections must be open

 Twist out the spindle pump counter-clockwise until it reaches the rear stop.

 Carefully pump using the initial pressure pump, while continuously observing the filling medium in

the open test connections. At this point, trapped air escapes toward the exterior by means of the
formation of bubbles. The initial pressure pump must be actuated until air bubbles no longer
appear.


 Any oil escaping in the open test connections should be siphoned off, for example, with a nozzle.