8 wet sieving, Wet sieving, Warning – Retsch AS 450 control Benutzerhandbuch

Seite 32: Notice

Advertising
background image

Wet sieving

32

8 Wet sieving

Pos : 7. 57 /00004 Warnhinweis e/ W0001 WARNUN G G ef ahr durc h Stromsc hl ag (AS450) @ 3\ mod_1282730212243_9.doc @ 22722 @ @ 1

WARNING

5.W0001

Danger from electric shock

An electric shock can result in burns and arrhythmia or to respiratory arrest
as well as cardiac arrest.

Never operate your AS450 in a water basin.

Always operate your AS450 using a mains socket with an RCD circuit
breaker.

Pos : 7. 58 /00004 Warnhinweis e/ W0009 WARNUN G Lebensg efahr durc h Str omsc hlag I P40 ( AS450) @ 3\ mod_1282730353954_9.doc @ 22730 @ @ 1

WARNING

6.W0009

Danger to life from electric shock

The control panel has safety class IP40.

An electric shock can result in burns and arrhythmia or to respiratory arrest
as well as cardiac arrest.

Ensure that the control panel does not come into contact with water.

Pos : 7. 59 /00004 Warnhinweis e/H 0048 HIN WEIS Nasssiebung ( AS450) @ 3\ mod_1282731343679_9.doc @ 22738 @ @ 1


NB

7.H0048

Wet sieving

Observe the following instructions for successful wet sieving :

The hose inner diameter of the water supply to the spray nozzle must be
13mm.

During the wet sieving a water pressure of between 2 bar and 5 bar is
recommended at the sieve cover entry (approx. 2 liters/min

– 4 liters/min).

These values depend on the sieve mesh size used, sieve material and
collecting pan loading

Use ventilation bases between the sieves.

The water outlet from the drain hose must be positioned below the water
sieve outlet (W1).

The distance between the water outlet of the drain hose and the water
sieve outlet should not be too great. (W2)

The inner diameter of the water drain hose must be sufficiently
dimensioned. (Hose inner diameter 20mm)

Used sieves must be cleaned immediately after sieving. Depending on the
sieved material, surface rust may develop in the sieve mesh.

Pos : 7. 60 /00004 Warnhinweis e/H 0049 HIN WEIS Besc hädigung das Si ebg ewebes (As 450) @ 3\ mod_1282731456436_9.doc @ 22746 @ @ 1

NOTICE

8.H0049

Damage to the sieve mesh

A build-up of water in the sieve column can lead to overloading and thus to
damage or irreparable damage to the sieve mesh.

Always dose the volume of water applied in a way that ensures there
is no build-up of water

Pos : 7. 61 /00004 Warnhinweis e/H 0050 HIN WEIS Belast ungs diagramme (As 450) @ 3\ mod_1282731525373_9.doc @ 22754 @ @ 1