Wet sieving – Retsch AS 450 control Benutzerhandbuch
Seite 34

Wet sieving
34
Fig. 18: Arrangement of the water system
NOTE
Wet sieving
–
For wet sieving to be successful, please observe the following instructions
:
•
The inside diameter of the water supply hose to the spraying nozzle must
be 13 mm.
•
During the wet sieving a water pressure of between 2 bar and 5 bar is
recommended at the sieve cover entry (approx. 2 litres/min
– 4 litres/min).
These values depend on the sieve mesh size used, sieve material and
collecting pan loading.
•
Use ventilation bases between the sieves.
•
The water outlet from the drain hose must be below the water sieve outlet
(W1).
•
The spacing between the water outlet from the drain hose and the water
sieve outlet should not be too large. (W2)
•
The inside diameter of the water drain hose must be sufficiently large.
(Inside diameter of hose is 20 mm)
•
Used sieves must be cleaned immediately after the sieving process.
Depending on the material being sieved, a rust film can form in the sieve
mesh.
8.1.2
Performing Wet Sieving
•
Add the solid material as a suspension .
W1
W2