Wet sieving – Retsch AS 450 control Benutzerhandbuch

Seite 34

Advertising
background image

Wet sieving

34

Fig. 18: Arrangement of the water system

NOTE

Wet sieving

For wet sieving to be successful, please observe the following instructions
:

The inside diameter of the water supply hose to the spraying nozzle must
be 13 mm.

During the wet sieving a water pressure of between 2 bar and 5 bar is
recommended at the sieve cover entry (approx. 2 litres/min

– 4 litres/min).

These values depend on the sieve mesh size used, sieve material and
collecting pan loading.

Use ventilation bases between the sieves.

The water outlet from the drain hose must be below the water sieve outlet
(W1).

The spacing between the water outlet from the drain hose and the water
sieve outlet should not be too large. (W2)

The inside diameter of the water drain hose must be sufficiently large.
(Inside diameter of hose is 20 mm)

Used sieves must be cleaned immediately after the sieving process.
Depending on the material being sieved, a rust film can form in the sieve
mesh.

8.1.2

Performing Wet Sieving

Add the solid material as a suspension .

W1

W2

Advertising