Wet sieving

34

Fig. 18: Arrangement of the water system

NOTE

Wet sieving

–

For wet sieving to be successful, please observe the following instructions
:

•

The inside diameter of the water supply hose to the spraying nozzle must
be 13 mm.

•

During the wet sieving a water pressure of between 2 bar and 5 bar is
recommended at the sieve cover entry (approx. 2 litres/min

– 4 litres/min).

These values depend on the sieve mesh size used, sieve material and
collecting pan loading.

•

Use ventilation bases between the sieves.

•

The water outlet from the drain hose must be below the water sieve outlet
(W1).

•

The spacing between the water outlet from the drain hose and the water
sieve outlet should not be too large. (W2)

•

The inside diameter of the water drain hose must be sufficiently large.
(Inside diameter of hose is 20 mm)

•

Used sieves must be cleaned immediately after the sieving process.
Depending on the material being sieved, a rust film can form in the sieve
mesh.

8.1.2

Performing Wet Sieving

•

Add the solid material as a suspension .

W1

W2