40

Whenthe

displayshows

Action/Description

Put the instrument in setup mode and scroll through the main menu

C Pn

Press ì or Å

C Py

Press

ENT to select input calibration

InLO

Set the signal source to the lowest value in the required range.
Wait a few seconds (for the signal to stabilise) and press

ENT

InLO

Display will stop flashing (for about 20 s) to show that the instrument is processing
the input.

InHI

Set the signal source to the Highest value in the required range.
Wait a few seconds (for the signal to stabilise) and press

ENT

InHI

Display will stop flashing (for about 10 s) to show that the instrument is processing
the input.

COPn

Continue with the setup sequence.

Note: To discard changes to calibration and setup you can remove the power at this point.