DACU_820_BA_V2_1M.doc

31/40

Baumer Electric AG

Frauenfeld, Switzerland

6.2.3

Operate, Reset

Turns the DACU on Reset resp. Operate.

Command data structure:

Start

command

Parameter

Checksum

02h

62h

31h

35h

STX

b

1

5 (95h)

Parameter

0:

Reset

1:

Operate

6.2.4

Fix Range

Setting a fix measuring range.

Command data structure
Example: CH2 10’000pC

Start

command

P. 1

P. 2

P. 3

Checksum

02h

63h

31h

30h

37h

44h

STX

c

1

0

7

D(FDh)

Parameter 1

0:

CH1

1

CH2

Parameter 2 & 3

02: Range 1

500'000 pC

CH1

03: Range 2

200'000 pC

CH1

04: Range 3

100'000 pC

CH1

06: Range 5

20'000 pC

CH2

07: Range 6

10'000 pC

CH2

08: Range 7

5'000 pC

CH2

09: Range 8

2'000 pC

CH2

6.2.5

Variable Range

Setting the variable range of CH1 between 100’000pC and 500’000pC

Command data structure:
Example: 230’000pC

Start command P. 1

P. 2

P. 3

P. 4

P. 5

P. 6

P. 7

Checksum

02h

64h

30h

32h

33h

30h

30h

30h

30h

42h

STX

d

0

2

3

0

0

0

0

B (1BBh)

Parameter 1

0:

CH1

Parameter 2 - 7

Numerical value ASCII-Format

DACU_820_BA_V2_1M.doc

32/40

Baumer Electric AG

Frauenfeld, Switzerland

6.2.6

Alarm outputs

Alarm output setting of the threshold.

Command data structure:
Example: Alarm 2 auf (0)8’500mV

Start

command

P. 1

P. 2

P. 3

P. 4

P. 5

P. 6

Checksum

02h

65h

31h

30h

38h

35h

30h

30h

35h

STX

e

1

0

8

5

0

0

5 (195h)

Parameter 1

0:

Alarm 1

1:

Alarm 2

Parameter 2 - 6

Numerical value in ASCII-Format

6.2.7

80% Test

Test function to check the measuring chain.
Output value 8V

Command data structure:
Example: Test on

Start

command

P. 1

Checksum

02h

66h

31h

39h

STX

f

1

9 (99h)

Parameter 1

0:

Test off

1:

Test on

6.2.8

Signal

Read out of the actual measuring value.

Command data structure:

Start

Command

P. 1

Checksum

02h

67h

31h

41h

STX

g

1

A (9Ah)

Parameter 1

0:

Read out of measuring value CH1

1:

Continuous read out of measuring values CH1

2

Read out of measuring value CH2

3

Continuous read out of measuring values CH2


Comment:

The continuous read out oft the measuring values can be stopped by changing the parameter 1
(Parameter 1 = 0 or 2)