IMPAC pyrometers IS 12-TSP · IGA 12-TSP

13

8

Instrument settings


All instrument settings can be done directly at the instrument. To activate the adjusting keys, a tip of ball pen
must be used. This avoids changing parameters by mistake.


8.1

Key panel operation

Temperature- or
parameter display






Targeting light on / off


Setting keys:

1) Parameter selection
2) Change of parameters:



3) Confirmation of parameter

Indication of limit switches
Indication °C or °F

Parameter indicator

Emissivity

Exposure time (t

90

/s)

Clear time of max. value storage (t

CL

/s)

Beginning of sub range (FROM)
End of sub range (TO)
Analog output 0 or 4 ... 20 mA (0/4 mA)
Setting of limit switches (SETP)
Temperature in °C or °F (°C/°F)
Interface RS232 or RS 485 (RS)
Address (ADR)
Baud rate (kBaud)
Test current output (10 mA)



8.2

Functions and setting of parameters

Temperature or
parameter display:

In the measuring mode the display shows the actual temperature reading. After

pushing the parameter key the display indicates the actual value of one selected
parameter.

Special indications: 8888 = measurement exceeds the end of sub range

Targeting light
on / off:

Pushing the PL key in measuring mode the laser targeting light is switched on.

The IS 12 measures continuously also with switched on targeting light and the
display shows the actual temperature reading. The laser targeting light is switched
off automatically after approx. 2 min or after a further push of the PL key.

Setting keys: 1) PAR:

2)



:


3)

ENT:

With the PAR button all available parameters are displayed in the following de-

scription. Pushing the button again changes the display to the next parameter,
after the last parameter it changes to the actual temperature reading.

With the arrow keys

and

all parameter settings can be displayed. Pushing

the button longer changes the settings in fast mode (the keys



and ENT are

blocked if the keyboard is locked via the interface commands or/and during a
measurement in combination with a PC and InfraWin).

If a parameter is changed with the arrow keys the new value must be confirmed

by pushing the ENT key. If it is not confirmed with ENT the instrument is still op-
erating with the previous parameter value. If no key is pressed for approx. 30 s
the display changes to the temperature indication.

Limit contacts:

Two limit switches can be set. If the measuring temperature exceeds the adjusted
limit contact temperature, the LED S1 or S2 displays the switch status.

Scale °C or °F:

The LED indicates the temperature scale in °C or °F.

Parameter indicator:

LED’s indicate which pyrometer parameter is selected for reading or changing.