1 dimensions, Dimensions – LumaSense Technologies IS 12-TSP Benutzerhandbuch

Seite 6

Advertising
background image

IMPAC pyrometers IS 12-TSP · IGA 12-TSP

6

address; baud rate; test current output

Additionally (only via interface adjustable):

key lock, wait time, recalibration (with special software)

Emissivity

:

0.100 to 1.000 in 1/1000 steps

Exposure time t

90

:

1 ms (with dynamical adaptation at low signal levels); factory setting 1 s,

adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s

Maximum value storage:

Built-in single or double storage. Clearing with adjusted time t

clear

(off; 0.01 s;

0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or automatically with the
next measuring object

Limit switches:

2 relay outputs (switch-over relay contacts), switch power max. 30 W
(I

max

: 1 A, U

max

: 60 V DC)

Rise and fall times: exposure time of the pyrometer + 2 ms

Control panel:

4 keys, operate with tip of ball-point pen

Protection class:

IP65 (DIN 40 050)

Ambient temperature:

0 to 70°C at housing

Storage temperature:

-20 to 70°C

Relative humidity:

Non condensing conditions

Weight:

2.2 kg

Sighting:

built-in parallax free thru-lens view finder, optionally
with an additional built-in laser targeting light (max.
power level < 1 mW,

 = 630-680 nm, CDRH class II)

CE-label:

According to EU directives about electromagnetic immunity



1.1

Dimensions


Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: