LumaSense Technologies IS 12-TSP Benutzerhandbuch
Seite 23
IMPAC pyrometers IS 12-TSP · IGA 12-TSP
23
Reading temperature
value repeated:
AAmsXXX
XXX = 000…999 (XXX = number of measuring values)
Reading basic tempera-
ture range:
AAmb
Output: XXXXYYYY (hex 8-digit, °C or °F)
XXXX = beginning of temp. range
YYYY = end of temp. range
Reading temperature
sub range:
AAme
Same as mb
Reading emissivity:
AAem
Output: XXXX (dec. 0010 … 1000 in ‰)
Emissivity:
AAemXX
XX = (10…99%), 00 = 100% (decimal)
Emissivity:
AAemXXXX
XXXX = (0010 … 1000‰) (decimal)
Exposure time t
90
:
AAezX
X = 0 ... 6 (decimal) 0 = intrinsic time constant of the device
1 = 0.01 s
3 = 0.25 s
5 = 3.00 s
2 = 0.05 s
4 = 1.00 s
6 = 10.00 s
External deletion:
AAlx
Simulation of an external deletion contact
Clear time of maximum
value storage:
AAlzX *
)
X = 0 ... 8 (dec.)
0 = Maximum value storage off
1 = 0.01 s
4 = 1.00 s
7 = external deletion
2 = 0.05 s
5 = 5.00 s
8 = automatic deletion
3 = 0.25 s
6 = 25.00 s
9 = hold
Setting of temperature
sub range:
AAm1XXXXYYYY XXXX (hex 4-digit) beginning of temp. range (°C)
YYYY (hex 4-digit) end of temp. range (°C)
Analog output:
AAasX *
)
X = 0 or 1 0 = 0...20 mA 1 = 4...20 mA
Limit contact 1:
AAs1XXXX
XXXX = switch point limit contact 1
(ASCII-hex, 4 digit, full degree in °C or °F)
Limit contact 2:
AAs2XXXX
XXXX = switch point limit contact 2
(ASCII-hex, 4 digit, full degree in °C or °F)
Setting hysteresis limit
contacts:
AAhlXX
XX = 2 … 20, full degree in °C or °F
Changing °C / °F
AAfhX *
)
Output: X = 0: display in °C; X = 1: display in °F
interface type:
AAin
Output: X = 1 or 2; 1 = RS232, 2 = RS485
Address:
AAgaXX *
)
XX = (00 ... 97)
00 ... 97 = regular device addresses
99 = global address with response
98 = global address without response (only setting commands!)
Baud rate:
AAbrX *
)
X = 1...6 or 8 (decimal)
1 = 2400 Baud
4 = 19200 Baud
(7 is not allowed)
2 = 4800 Baud
5 = 38400 Baud
8 = 115200 Baud
3 = 9600 Baud
6 = 57600 Baud
Wait time:
AAtwXX *
)
XX = 00 ... 99 (decimal)
Error status:
AAfs
Output: XX; XX=00…FF (00 = no error)
(01…FF: error code for LumaSense service)
Lock keyboard:
AAlkX
X = 0 ... 3
1 = lock lk1, removal with command lk0 or power off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command lk2
2 = removal of lock lk3
Reading parameters:
AApa
Output decimal 11-digit:
Digit 1 und 2 (10...99 or 00):
(emissivity)
Digit 3 (0 ... 6): t
90
(exposure time)
Digit 4 (0 ... 8): t
CL
(max. value storage clear mode)
Digit 5 (0 ... 1): analog output
Digit 6 und 7: (00 ... 98): temperature
Digit 8 und 9 (00 ... 97): address
Digit 10 (0 ... 6 or 8): baud rate
Digit 11 (0): always 0
Laser targeting light:
AAlaX *
)
X = 0 switch off laser
X = 1 switch on laser
Internal temperature:
AAgt
Output: XXX (dec. 000 ... 098°C or 032 ... 208°F)
Max. internal temp.:
AAtm
Output: XXX (dec. 000 ... 098°C or 032 ... 208°F)