IMPAC pyrometers IS 12-TSP · IGA 12-TSP

5

Legend

Note: The note symbol indicates tips and useful information in this manual.
All notes should be read with regard to an effective operation of the instrument.

Security note laser beam
Indicates to the danger of a built-in laser targeting light.

MB

Shortcut for Temperature range (in German: Messbereich)


Terminology

The used terminology corresponds to the VDI- / VDE-directives 3511, page 4.

Disposal / decommissioning

Inoperable IMPAC pyrometers have to be disposed corresponding to the local regulations of electro or elec-
tronic material.

1

Technical data

Temperature ranges:

IS 12-TSP:

600 to 2520°C IGA

12-TSP:

200 to 1020°C

600 to 3000°C

250 to 1400°C

850 to 2520°C

Spectral range:

IS 12-TSP:

0.94 µm (at temperature range 600 to 2520°C / 3000°C)

0.65 µm (at temperature range 850 to 2520°C)

IGA 12-TSP: 1.57 µm

Accuracy

1

(

 = 1, t

90

= 1 s, T

amb.

= 23°C)

Below 1500°C: 0.15% of measured value in °C + 1°C
Above 1500°C: 0.25% of measured value in °C
Above 2700°C: 0.35% of measured value in °C

Repeatability:

(

 = 1, t

90

= 1 s, T

amb.

= 23°C)

1°C

Resolution:

Up to 1000°C:

0.1°C at interface and display

Above 1000°C: 0.1°C at display, 0.01°C at interface
< 0.025% of temperature range at the analog output

Power supply:

24 V DC (15 to 40 V DC) or 24 V AC (12 to 30 V AC), 48 to 62Hz

Power consumption:

max. 14 W

Analog output:

0 to 20 mA or 4 to 20 mA switchable

Load:

0 to 500 Ohm

Test current output:

10 mA, fixed

Digital interface:

Switchable at the instrument: RS232 or RS485 addressable (half duplex),
baud rate 2.4 up to 115.2 kBd

Display:

Built-in 5 digit LED display, height 13 mm, additional function LED’s

Isolation:

Power supply, digital interface, analog output are galvanically isolated
against each other and housing

Sub range:

Any range adjustable within the temperature range
minimum span 51°C

Signal processing:

Photoelectric current, digitized

Parameters:

Adjustable at the instrument or via serial interface:

emissivity

; response time t

90

; clear times t

CL

of the maximum value sto-

rage; temperature sub range; analog output 0 to 20 or 4 to 20 mA;
switch points for limit contacts; °C or °F; interface RS232 or RS485;

1

Note: A surrounding temperature of 10 to 43°C is permissible for measured temperatures between 10%

and 100% of the temperature range for the stated measurement uncertainty. The measurement uncertainty
increases by 1°C for surrounding temperatures of between 0 to 10°C or 43 to 60°C.
For temperature ranges of up to 5% of the measuring range an additional temperature drift of 0.2°C/°C
(23°C) should be added. The temperature drift is 0.05°C/°C (23°C) for measured temperatures of 5% to
10% of the measuring range.