1 emissivities of common materials, 6 factory settings, 8 software settings – LumaSense Technologies IGA 5-LO Benutzerhandbuch

Seite 16: Emissivities of common materials, Factory settings, Software settings, Clear time intervals t, Integrated maximum value storage), Cted (see 8.1 clear tim

Advertising
background image

IMPAC pyrometers IS 5 · IS 5-LO · IGA 5 · IGA 5-LO

16

7.5.1

Emissivities of common materials

Emissivity [%]

Emissivity [%]

Measuring object

IS 5

IS 5-LO

(0.8...1.1 µm)

IGA 5

IGA 5-LO

(1.45...1.8 µm)

Measuring object

IS 5

IS 5-LO

(0.8...1.1 µm)

IGA 5

IGA 5-LO

(1.45...1.8 µm)

„Black body furnace“

100

100

Zinc

58

45 to 55

Steel heavily scaled

93

85 to 90

Nickel

22

15 to 20

Steel rolling skin

88

80 t 88

Gold, Silver, bright

2

2

Steel, molten

30

20 to 25

Porcelain glazed

60

60

Slag

85

80 to 85

Porcelain rough

80 to 90

80 to 90

Aluminum, bright

15

10

Graphite

80 to 92

80 to 90

Chromium, bright

28 to 32

25 to 30

Chamotte

45 to 60

45 to 60

Brass oxidized (tarnished) 65 to 75

60 to 70

Earthenware, glazed

86 to 90

80 to 90

Bronce, bright

3

3

Brick

85 to 90

80 to 90

Copper, oxidized

88

70 to 85

Soot

95

95

Set up examples at the pyrometer:

Emi =100%:

Emi = 85%:

Note:

If the settings directly at the pyrometer should be used the pyrometer must be
switched into offline mode

Note:

If the emissivity is set to a value below 20% the instrument will automatically utilize
an emissivity value of 20%. The setting 00 is interpreted as

ε

= 100%!



7.6

Factory settings

Emissivity

ε

= 100%

Exposure time t

90

= 0.00 (corresponds to the instrument’s min. exposure time of ≤ 2 ms)

Analog output = 0 to 20 mA
Operating mode = online
Wait time (for RS485) = 03 Bit

8

Software settings


The digital PC interface enables the exchange of data with a PC either by using the InfraWin software (sup-
plied) or an user adapted communication software (see command table in section 13, Data format UPP).
There are also several additional detailed navigation/reading options. The descriptions to the settings for
emissivity

ε

, exposure time t

90

and analog output 0 or 4 to 20 mA are the same as described already in section

7

, Instrument settings. The operating switches must be switched to the online position before performing the

setting adjustments “emissivity”, “exposure time” and “analog output” with the software. The following additional
setting /reading options are available.


8.1

Clear time intervals t

cl

(integrated maximum value storage)

If the maximum value storage is switched on always the highest last temperature value
will be displayed and stored. The storage has to be cleared at regular intervals for ex-
changing by a new and actual value.
This feature is particularly useful when fluctuating object temperatures cause the display
or the analog outputs to change too rapidly, or the pyrometer is not constantly viewing an
object to be measured. In addition, it may also be beneficial to periodically delete and
reset the stored maximum values.

The following settings are possible:

OFF:

At clear time “OFF“ the max. value storage is switched off and only momentary values are
measured.

Settings:

OFF

0.01 s

. . .

25 s

extern

auto

Hold

Advertising