Legend, Terminology, Disposal / decommissioning – LumaSense Technologies IGA 5-LO Benutzerhandbuch

Seite 5: 1 technical data, Technical data

Advertising
background image

IMPAC pyrometers IS 5 · IS 5-LO · IGA 5 · IGA 5-LO

5

Legend

Note:

The note symbol indicates tips and useful information in this manual.

All notes should be read with regard to an effective operation of the instrument.

Security note laser beam
Indicates to the danger of a built-in laser targeting light.

MB

Shortcut for Temperature range (in German: Messbereich)

Terminology

The used terminology corresponds to the VDI- / VDE-directives 3511, page 4.

Disposal / decommissioning

Inoperable IMPAC pyrometers have to be disposed corresponding to the local regulations of electro or elec-
tronic material.

1

Technical data

Temperature range:

IS 5

:

600 to 2000°C (MB 20)

IGA 5

:

350 to 1800°C (MB 18)

800 to 2500°C (MB 25)

250 to 2000°C (MB 20)

1000 to 3000°C (MB 30)

400 to 2500°C (MB 25)

500 to 3000°C (MB 30)

IS 5-LO

: 600 to 2000°C (MB 20)

IGA 5-LO

: 300 to 1300°C (MB 13)

800 to 2500°C (MB 25)

350 to 1800°C (MB 18)

1000 to 3000°C (MB 30)

400 to 2500°C (MB 25)

500 to 3000°C (MB 30)

Sub range:

User adjustable (minimum span is 51°C)

Data handling:

digital

Spectral response

IS 5; IS 5-LO:

0.8 to 1.1 µm

IGA 5; IGA 5-LO: 1.45 to 1.8 µm

IR detector:

Silicon photo diode (Si); Indium-Gallium-Arsenic photo diode (InGaAs)

Power supply:

24 V DC ± 25% nominal, ripple must be less than 50 mV
maximum current requirement: < 3 VA (incl. active laser targeting light)

Analog output:

0 to 20 mA or 4 to 20 mA, linear, switchable

Load:

0 to 500

Ω

Digital interface:

RS232 or RS485 addressable (half-duplex),
baud rate 1200 to 38400 Bd, resolution < 0.1°C

Resolution:

0.1°C on interface
< 0.1% of the adjusted sub range on the analog output

Isolation:

Power supply, analog output and digital interface are galvanically
isolated from each other

Parameters:

Adjustable on the pyrometer:

Emissivity (

ε), exposure time t

90

, analog output 0 to 20 or 4 to 20 mA,

online-/offline-switch for adjustments at pyrometer or PC

Additionally adjustable or readable via digital interface:

Sub-temperature range (within the full measuring range), external
clearing of the maximum temperature value, clear times of the max.
value storage, hold function, address, baud bate, internal temperature.

pyrometers with PID-controller: set point, proportional band, rate time

/integral time, output delimitation (adjustable via software)

Emissivity:

20 to 100% adjustable in the instrument or

via serial interface in steps of 1%

Maximum value store

Built-in single or double storage. Clearing with adjusted time t

clear

(OFF; 0.01 s;

0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or automatically with the
next measuring object

Exposure time t

90

:

2 ms; adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s

Advertising