14 reference numbers, 1 reference numbers of instruments, Reference numbers – LumaSense Technologies IGA 5-LO Benutzerhandbuch

Seite 28: Reference numbers of instruments

Advertising
background image

IMPAC pyrometers IS 5 · IS 5-LO · IGA 5 · IGA 5-LO

28


AARsX

Bit 0 = 1

gain

∞ Æ on/off controller

X = 0 or 1 Æ deactivates / activates the controller

starting self-tuning:

AARt
AARtXXXX

Starts self-tuning algorithms
XXXX new set point (comp. Xs) and self-tuning-start

Note:

PID-controller: at °F setting

X

s

Desired temperature in °F x 10

R

p

Desired temperature output is °F x 10

Y

i

Reading controller output and actual value,

R

t

Self tuning with take over of controller output in °F x 10

actual value in °F x 10

Note: the letter ”l” means the small type of “L”


Additional instruction for the RS485 interface:
Requirements to the master system during half-duplex operation:

1. After an inquiry, the bus should be switched into a transmission time of 3 bits

(some older interfaces are not fast enough for this).

2. The pyrometer's response will follow after 5 ms at the latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be repeated.
After receiving the response, the master has to wait at least 1.5 ms before a new command can be entered.

14

Reference numbers


14.1

Reference numbers of instruments

IS 5; IGA 5

Type

Temperature range

targeting

light, RS232

view finder,

RS232

targeting

light, RS485

view finder,

RS485

IS 5

MB 20:

600 to 2000°C

3 857 150

3 857 170

3 857 160

3 857 180

IS 5

MB 25:

800 to 2500°C

3 857 200

3 857 220

3 857 210

3 857 230

IS 5

MB 30: 1000 to 3000°C

3 857 250

3 857 270

3 857 260

3 857 280

IGA 5

MB 18:

350 to 1800°C

3 857 400

3 857 420

3 857 410

3 857 430

IGA 5

MB 20:

250 to 2000°C

3 857 350

3 857 370

3 857 360

3 857 380

IGA 5

MB 25:

400 to 2500°C

3 857 450

3 857 470

3 857 460

3 857 480

IGA 5

MB 30:

500 to 3000°C

3 857 920

3 857 940

3 857 930

3 857 950


IS 5-C; IGA 5-C (with PID controller)

Type

Temperature range

targeting

light, RS232

view finder,

RS232

targeting

light, RS485

view finder,

RS485

IS 5-C

MB 20:

600 to 2000°C

3 851 150

3 851 170

3 851 160

3 851 180

IS 5-C

MB 25:

800 to 2500°C

3 851 200

3 851 220

3 851 210

3 851 230

IS 5-C

MB 30: 1000 to 3000°C

3 851 250

3 851 270

3 851 260

3 851 280

IGA 5-C

MB 18:

350 to 1800°C

3 851 400

3 851 420

3 851 410

3 851 430

IGA 5-C

MB 20:

250 to 2000°C

3 851 350

3 851 370

3 851 360

3 851 380

IGA 5-C

MB 25:

400 to 2500°C

3 851 450

3 851 470

3 851 460

3 851 480

IGA 5-C

MB 30:

500 to 3000°C

3 851 920

3 851 940

3 851 930

3 851 950


IS 5-TV; IGA 5-TV (with video module)

Type

Temperature range

RS232

RS485

IS 5-TV

MB 20:

600 to 2000°C

3 847 150

3 847 160

IS 5-TV

MB 25:

800 to 2500°C

3 847 200

3 847 210

IS 5-TV

MB 30: 1000 to 3000°C

3 847 250

3 847 260

IGA 5-TV

MB 18:

350 to 1800°C

3 847 400

3 847 410

IGA 5-TV

MB 20:

250 to 2000°C

3 847 350

3 847 360

IGA 5-TV

MB 25:

400 to 2500°C

3 847 450

3 847 460

IGA 5-TV

MB 30:

500 to 3000°C

3 847 920

3 847 930




Advertising