3.2 Remote control commands

795 KFT Titrino

69

3.2.2.11. Mode.Parameter.CtrlPara.EP Ipol:

0...250...±2000

Upol:

0...25...±200.0

Mode.Parameter.CtrlPara.UnitEp read

only

Setting of the EP, resp. Control point in mV (with Ipol) or µA (with Upol). The

corresponding unit can be read with .UnitEP.

3.2.2.12. Mode.Parameter.CtrlPara.Dyn Ipol: 1...100...2000

Upol:

0.1...10...200.0

Mode.Parameter.CtrlPara.UnitDyn read

only

Mode.Parameter.CtrlPara.MaxRate 0.01...10...150, max.

Mode.Parameter.CtrlPara.MinIncr 0.1...9.9, min.

Control parameters.
.Dyn: Dynamics (control range) in mV (with Ipol) or µA (with Upol). The cor-

responding unit can be read with .UnitDyn.

.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum

possible rate with mounted Exchange Unit.

.MinIncr: Minimum volume increment in µL.

OFF means that the criterion is not monitored.

3.2.2.13. Mode.Parameter.CtrlPara.Stop.Type drift, time

Mode.Parameter.CtrlPara.Stop.Drift 1...20...999

Mode.Parameter.CtrlPara.Stop.Time 0...10...999, inf.

Mode.Parameter.CtrlPara.Stop.StopT 0...99999, OFF

Type and size of the stop criterion of the titration.

.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in µL/min. Applies when "drift" has been selected.

.Time: Switch-off delay time in s. Applies when "time" has been selected.

"inf." means infinite.

.StopT: Stop time in s. Applies when "time" has been selected and the value of

.Time is set to "inf.".

3.2.2.14. Mode.Parameter.TitrPara.Direction +,

, auto

Titration direction.

"auto" means the titration direction is determined automatically by the instru-

ment.

3.2.2.15. Mode.Parameter.TitrPara.XPause

0...999999

Pause time in s. Runs before dosing the start volume.