3.2 Remote control commands

795 KFT Titrino

71

3.2.2.22. Mode.Parameter.StopCond.VStop.Type abs., rel., OFF

Mode.Parameter.StopCond.VStop.V 0...99.99...9999.99

Mode.Parameter.StopCond.VStop.Factor 0...±999999

Stop volume.

If an absolute stop volume (abs.) has been selected, the volume in mL is

valid.

A relative stop volume (rel.) is dispensed as a function of the sample size:
Stop volume in mL = smpl size ∗ factor

The factor is valid.

OFF means that the criterion is not monitored.

3.2.2.23. Mode.Parameter.StopCond.FillRate 0.01...150, max.

Filling rate in the titration in mL/min. Max. means maximum possible filling

rate with the Exchange Unit in current use.

3.2.2.24. Mode.Parameter.Statistics.Status ON, OFF

Mode.Parameter.Statistics.MeanN

2...20

Mode.Parameter.Statistics.ResTab.Selected original, delete n,

delete

all

Mode.Parameter.Statistics.ResTab.DelN

1...20

Entries for the statistics calculations.

.Status:

On/off switching. Requirement for statistics calculations is a valid

assignment, see 3.2.2.9.

.MeanN: Number of individual results for statistics calculations.

.ResTab.Select: Selection of the table for the statistics calculations.

original: Original table. The original table is (again) set up, i.e. any

individual results which have been deleted are reincorporated

in the statistics calculations.

delete n: Single result lines are removed from the statistics calcu-

lation. All results of the corresponding line in the statistics

table are deleted. Specification of the line number in

.ResTab.DelN.

delete all: Clear entire statistics table. The results can not be reac-

tivated.

.ResTab.DelN: Specification of the line number to be deleted.

3.2.2.25. Mode.Parameter.Presel.Cond

ON, OFF

Mode.Parameter.Presel.DriftDisp ON, OFF

Mode.Parameter.Presel.DCor.Type auto,

man., OFF

Mode.Parameter.Presel.DCor.Value 0.0...99.9

.Cond: Conditioning ON/OFF

.DriftDisp: Drift display during conditioning ON/OFF.

.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of

the drift value at start.

.DCor.Value: Drift value for the manual drift correction.