KROHNE CORIMASS MFC 85 DE Benutzerhandbuch
Seite 14

14
Enter the raw data (bold) into the table below then for each temperature calculate
ρ
c
and
ρ
s
using the above equations.
Temp. /°C
Conc. by Mass /%
Density
/g/cm
3
ρ
c
ρ
s
10
5.52
1.040473
1.002565
2.9473856
14.9
1.112023
20
5.52
1.037835
1.0006212
2.8540898
14.9
1.107953
30
5.52
1.034454
0.9977574
2.7906384
14.9
1.103516
Transfer the
ρ
c
and
ρ
s
into the table below and calculate the
ρ
20
, K1 and K2 values for each
using the equations of Case 2:
Density
Temp.
K
2
K
1
ρ
20
Liquid
ρ
1
= 1.002565
10
-0.0000046
-0.0002404
1.000621
ρ
2
= 1.0006212
20
ρ
3
= 0.9977574
30
Solid
ρ
1
= 2.9473856
10
+0.0001492
-0.0078374
2.85409
ρ
2
= 2.8540898
20
ρ
3
= 2.7906384
30
Program these values into the converter. For this example the fit of the concentration
algorithm, over the range 3 to 17%, is accurate to better than
±
0.1
4.4
Parametereingabe in den Meßumformer
Wenn die erforderlichen Kennwerte errechnet sind, müssen sie noch in den Meßumformer
eingegeben werden. Bei der Berechnung von
ρ
20
, K
1
, K
2
ist die Maßeinheit für die Dichte
(g/cm
3
, kg/m
3
, lb/US Gallon usw.) unerheblich, jedoch müssen diese Maßeinheiten dann für
alle Dateneingaben benutzt werden.
Betrachten wir beispielsweise eine Emulsion aus Öl und Wasser, in der die Öldichten 650.0,
648.5 und 647.3 kg/m
3
bei Temperaturen von 10, 20 und 30°C betragen. Aus diesen Werten
ergibt sich:
ρ
20
=
648.5 kg/m
3
K
1
=
-0.135 kg/m
3
/°C
K
2
=
0.0015 kg/m
3
/°C
2