Error messages

HART

®

communications errors are reported with the numerical codes 1 - 9. An error

message occurs when a transaction is not successful, despite being retried several times.

An error message is only displayed when an error occurs as a consequence of a command

sent from the display, i.e. only when an operation by the user leads to the error. Errors in the

communication between the control system and sensor are detected but not displayed.

The measuring value line shows "ERROR", the status line shows "HART

®

n" whereby "n"

represents the numeric error code. The error message remains displayed until the next key

actuation takes place.

The following table shows the error codes and information regarding the failure cause and

elimination.

Error code

Description

Causes

1

Transmitter does not respond

2

Communication error

Parity, check sum, package length incorrect

upon receiving

Transmitter signals receive error

3

Command not implemented

Command not supported by the transmitter

4

Range error

The value is outside the range supported by the

sensor

The desired unit of measure is not supported

5

Unspecified error of the

transmitter

The sensor signals an error in the status byte that

is not broken down

CAUTION!

If faults cannot be eliminated by means of the measures listed above, the

instrument must be shut down immediately, and it must be ensured that signal

is no longer present, and it must be prevented from being inadvertently put back

into service. In this case, contact the manufacturer.

If a return is needed, follow the instructions given in chapter 12.1 "Return".