WIKA TR10-B Benutzerhandbuch

Seite 12

Advertising
background image

3132465.01 07/2011 GB/D

12

WIKA operating instructions resistance thermometers per EN 14597

GB

Store the instrument in its original packaging in a location that fulfils the conditions listed

above. If the original packaging is not available, pack and store the instrument as described

below:

1. Wrap the instrument in an antistatic plastic film.

2. Place the instrument, along with shock-absorbent material, in the packaging.

3. If stored for a prolonged period of time (more than 30 days), place a bag containing a

desiccant inside the packaging.

WARNING!

Before storing the instrument (following operation), remove any residual media.

This is of particular importance if the medium is hazardous to health, e.g. caustic,

toxic, carcinogenic, radioactive, etc.

6. Commissioning, operation

WARNING!

When the thermometer is mounted, the temperature must not drop below the

allowed operating temperature (environment, medium) or exceed it, even when

taking convection and heat radiation into account!

WARNING!

Thermometers must be earthed if dangerous voltages could be expected at the

connection wires (caused, for example, by mechanical damage, electrostatic

discharge or induction)!

6.1 Electrical connection

CAUTION!

Damage to cables and wires, and to connection points must be avoided

Flying leads with bare ends must be finished with ferrule ends (cable

preparation)