WIKA operating instructions resistance thermometers per EN 14597

3132465.01 07/2011 GB/D

9

GB

3. Specifications

3.1 Resistance thermometer
Sensor connection method

2-wire The lead resistance is recorded as an error in the measurement.

3-wire With a cable length of approx. 30 m or longer, measuring errors can occur.

4-wire The internal lead resistance of the connecting wires is negligible.

Sensor tolerance value per DIN EN 60751

Class B

Class A

Class AA

The combinations of a 2-wire connection with class A or class AA are not permissible, since

the lead resistance of the measuring insert negates the higher sensor accuracy.

Basic values and limiting errors

Basic values and limiting errors for the platinum measurement resistances are laid down in

DIN EN 60751.

The nominal value of Pt100 sensors is 100 Ω at 0 °C.

The temperature coefficient α can be stated simply to be between 0 °C and 100 °C with:

α = 3,85 ∙ 10

-3

°C

-1

The relationship between temperature and electrical resistance is described by

polynomialss, which are also defined in DIN EN 60751. Moreover, this standard specifies the

basic values in °C steps in tabular form.

Class

Temperature range

Tolerance value in °C

Wire-wound (W)

Thin-film (F)

B

-196 … +600 °C

-50 … +500 °C

±(0.30 + 0.0050 | t |)

1)

A

-100 … +450 °C

-30 … +300 °C

±(0.15 + 0.0020 | t |)

1)

AA

-50 … +250 °C

0 … 150 °C

±(0.10 + 0.0017 | t |)

1)

1) | t | is the value of the temperature in °C without consideration of the sign.

Bold: Standard version