Gb 6. commissioning, operation – WIKA CPG1000 Benutzerhandbuch

Seite 25

Advertising
background image

WIKA operating instructions model CPG1000

25

11568047.02 08/2012 GB/D

GB

6. Commissioning, operation

Pressure units

The instrument is factory configured to display "PSI". Using the ▲ and ▼ keys (ZERO

and MAX/MIN) you can scroll up and down through the 18 engineering units defined as a

standard, and through the user-defined units or scale. Once the required measuring unit

is displayed, press the Enter key or the CONFIG key. The pressure will then be displayed

in the selected engineering units. A list of available engineering units is given in chapter

3 "Specifications", of this manual. Details on the configuration of user-defined measuring

units is given in chapter 6.3 "Supervisor mode".

Automatic Power-off

The function for automatic shut-down can be set in intervals of 1 minute, from 1 to

30 minutes, or it can be turned off (continuous operation). The measuring instrument is

delivered with an automatic power-off after 30 minutes. Define the required interval using

the ▲ and ▼ keys. The OFF setting is at the very start of the options before setting for

1 minute.

Displaying battery voltage

The current battery voltage, and the remaining useful life of the battery, are displayed in the

form of a progress bar display.

Indication of the current temperature

The measuring instrument is temperature compensated. This option displays the tempera-

ture measured by the internal sensor. The display value can be switched from degrees

Fahrenheit to degrees Celsius.

Setting damping

Using the ▲ and ▼ keys you can turn this option on and off. With damping enabled, the

measured values from pulsating pressure sources are smoothed.

Configuring the sampling rate

The sampling rate defines the frequency with which the pressure is measured and the

indication is refreshed. The options are 0.5, 1, 3 and 10 measurements per second. The

shortest response time is 10 measurements per second.

TARE (setting Tare)

Using this option you can set the constant offset value which will be subtracted from the

measured pressure. If, for example, TARE is set to 30 bar and the measured pressure is

37 bar, the measured value displayed will be 7 bar. A pressure of 27 bar would then be

displayed as -3 bar. The TARE value is set manually using the ▲ and ▼ keys, it depends on

the engineering units and the resolution selected for the indication. The TARE value can be

set to the maximum scale range. The progress bar indication always indicates the actual

pressure in relation to the entire measuring range, irrespective of the Tare setting. This is

performed for safety reasons so that when "0" is displayed, it can still be shown that there

is a corresponding pressure at the pressure gauge.

Advertising