Gb 8. faults 8. faults, Indication causes measures – WIKA CPG1000 Benutzerhandbuch

Seite 28

Advertising
background image

28

WIKA operating instructions model CPG1000

11568047.02 08/2012 GB/D

GB

8. Faults

8. Faults

Indication

Causes

Measures

Battery is low, functioning is only

guaranteed for a short period of

time

Insert new alkaline batteries

OL

-OL

Reading is significantly above or

below the measuring range

Test: if pressure is within the

approved measuring range of the

sensor?

No display or un-

definable charac-

ters, the instrument

does not respond

to key press

Battery is dead.

Insert new alkaline batteries

Operation of mains adapter: false

voltage/polarity

Check/replace mains adapter

System fault

Turn off the CPG1000, wait for

a short period of time, switch on

again

CPG1000 is defective

Send in for repair

CAUTION!

If faults cannot be eliminated by means of the measures listed above, the

CPG1000 must be shut down immediately, and it must be ensured that

pressure and/or signal are no longer present, and it must be prevented

from being inadvertently put back into service.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter

9.2 "Return".