WIKA operating instructions model CPG1000

11568047.02 08/2012 GB/D

GB

2. Safety

Further hazardous area safety instructions:

Observe the operating information and the relevant country-specific regula-

tions concerning use in hazardous areas (e.g. EN 60079-14).

The precision digital pressure gauge is not intended for use with flammable

substances. It is only suitable for installation in locations which provide

sufficient protection against the intrusion of solid foreign objects or water

which could impair the safety.

The instructions in these operating instructions must be followed strictly in

order to guarantee safe operation.

Information/Approval for hazardous locations

WARNING!

Hazardous areas

An hazardous area, as used in this manual, refers to an area made

hazardous by the potential presence of flammable or explosive vapours.

These areas are also referred to as hazardous locations, see NFPA 70

Article 500.

® LR110460, Class I, Division 2, Group A, B, C, D

II 3G Ex nA IIB T6

KEMA 09ATEX0096 X

T

a

= -10 ... +55 °C

WARNING!

Only precision digital gauges powered by alkaline batteries are approved

for use in hazardous rooms. Precision digital pressure gauges with an

external voltage supply are not approved for use in hazardous areas.

The replacement of any components impairs their suitability for use in

hazardous areas.