WIKA Betriebsanleitung Hand-Held Thermometer

5

GB

11044608.02 07/2006 D/GB

WIKA Operating Instructions Test Pump

Test Pump
CPP30

3. Mounting Instructions

The reference instrument is fitted to the upper side of the hand
test pump CPP30. Fingertight fastening of the reference instru-
ment with the knurled nut is sufficient. The reference instrument
is sealed by the integral O-ring sealing gasket.

The unit under test is mounted to the end of the flexible tube.
Please use a suitable sealing gasket from the adapter set
purchased separately as an accessory. Tighten the connector to
prevent any leaks to a maximum torque of 15 Nm.
In order to adapt the different connection threads of the unit
under test, the test tube can be fitted with different adapters
from the optional set of adapters.

Do not use teflon tape, this may damage your test pump.

You can unscrew the tube and also directly attach the test
specimen with the same adapter to the pump (to minimize
volume of your test system, for more easy operating the
CPP30 pump).