Correct use, Unpacking – IKA LR 1000 basic Package Benutzerhandbuch

Seite 24

Advertising
background image

24

Before you fill the reactor vessel, en-
sure that the reagents used do not
corrode the FTFE and FKM seal.

• Ensure that the external temperature sensor is inserted in

the media to a depth of at least 20 mm.

• The external temperature sensor must always be inserted

in the media when connected.

• Only use IKA

®

approved accessories!

• Use only original IKA

®

spare parts!

• When the reactor vessel or the vessel cover is removed

from the LR 1000 base during operation, the stirring and

the heating function will be switched off automatically.
Restart the stirring and heating function by pressing on
knob (A) and (B) after the reactor vessel or vessel cover is

placed and locked again.

Weighing function integrated into
the LR 1000 conttol feet. Unpack

and set up cautiously.

Risk of burns! Exercise caution when
touching the heating block and the
glass vessel. The heating block can

reach temperatures in excess of 230 °C and the glass ves-
sel can reach 120 °C. The cover of the vessel could reach
high temperature when you boil liquid for long time. Pay
attention to the residual heat after switching off.

• Ensure that the thermostat used for tempering is fully ef-

ficient. A defective thermostat can cause uncontrolled
reactions.

Only process and heat up any media
that has a flash point higher than
the adjusted safe temperature limit

that has been set (50...210 ºC). The safe temperature limit
must always be set to at least 25 °C lower than the fire
point of the media used.

• The equipment may heat up by stirring.
• Gradually increase the speed.

Use

The IKA

®

LR 1000 system is a compact reactor system.

It has been designed to simulate and optimize chemical
reaction processes and can also be used for mixing, dis-
persion and homogenization in modeling processes.

Operating mode: Tabletop device.

• Range of use (indoor use only)

- Laboratories

- Schools

- Pharmacies

- Universities

Correct use

This equipment is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply

network that also supplies residential areas.

The safety of the user cannot be guaranteed:
- if the instrument is operated with accessories that are not

supplied or recommended by the manufacturer

- if the instrument is operated improperly or contrary to the

manufacturer’s specifications

- if the instrument or the printed circuit board are modified

by third parties.

DANGER

DANGER

WARNING

• Unpacking

- Please unpack the device carefully
- In the case of any damage a detailed report must be

sent immediately (post, rail or forwarder).

• Delivery scope

- LR 1000 basic or LR 1000 control base according

to order

- LR 1000 reactor vessel

- LR 1000.61 sensor receptacle

- Support rod
- 2 hose connectors
- Temperature sensor Pt 100.5

- USB cable (only for LR 1000 control)

- Mains cable
- Tools kit

see Fig. 1

- Operating instructions
- Warranty card.

Unpacking

Fig. 1

WARNING

Advertising