Attach the reactor vessel to the LR 1000 basic/control base

Ensure the LR 1000 basic/control is set up on an even,

stable, clean and non-slip working table.
Set the reactor vessel to the LR 1000 basic/control base

carefully and ensure it is properly attached as indicated in
following image.

Setting up

Installation and connection Pt 100 temperature sensor

Remove a standard connection NS 14/23 from the reactor
cover for connection the temperature probe.
Screw the LR 1000.61 sensor receptacle with sealing in

the reactor cover.
Plug in the probe and pay attention to the minimum immer-
sion depth and freewheeling of the anchor mixer. Fix the posi-
tion in the housing with the two screws for sliding stop.
Connect the temperature probe with the appropriate socket.

Connect external cooler

The reactor can connect an external cooler (e.g. IKA

®

KV 600) via the connections (13, 14, see Fig. 3) at the

back of the device. The cooling connections IN and OUT is

labelled accordingly at the back of the device. There are two
connectors included for connecting cooling hoses. They can
be connected to a hose with a 10 mm inner diameter. By
positioning and light pressure in the direction of the con-
nector, the connector is locked/connected to the cooling
connections IN/OUT on the housing.

The connectors are unlocked by coaxially positioning the
unlocking lever and pressing slightly.

CLICK

Unlock

Lock

Fig. 6

Fig. 8

Attach the support rod to the LR 1000 basic/control base

Screw the support rod onto the LR 1000 basic/control

base with the double open end wrench included with the
device until the end stop is reached.

Fig. 5

Fig. 7

Pt 100