IKA LR 1000 basic Package Benutzerhandbuch

Seite 35

Advertising
background image

35

Abbreviations used:
X,y = numbering parameter (integer number)
m =

variable value, integer

n =

value of variable, floating point number

X = 1

Pt100 thermometer (external temperature sensor)

X = 2 temperature (heating block)
X = 3 safety temperature
X = 4 stirring speed
X = 6 safety stirring speed
Commands

NAMUR Commands

Function

IN_NAME

Input description name

IN_PV_X
X=1;2;3;4;

Reading the real value

IN_SOFTWARE

Input software ID number date, version

IN_SP_X
X=1;2;3;4;6;

Reading the set rated value

IN_TYPE

Input laboratory unit ID

OUT_NAME

Output description name. (Max. 10 characters, default: LR 1000)

OUT_SP_12@n

Setting the WD safety temperature with the echo of the set value

OUT_SP_42@n

Setting the WD safety speed with the echo of the set value

OUT_SP_X n
X=1;2;4;6

Setting the rated value to n

OUT_WD1@m

Watchdog mode 1: When a WD1 event occurs, the heating and shaking functions are
shutdown and message PC 1 is displayed. Set the watchdog time to m (20...1500) seconds,
with echo of the watchdog time. This instruction starts the watchdog function and must be
sent within the set watchdog time.

OUT_WD2@m

Watchdog mode 2: When a WD2 event occurs, the speed setpoint will be set to the WD
safety setpoint speed and the temperature setpoint will be set to the WD safety setpoint
temperature. The PC 2 warning is displayed. The WD2 event can be reset with OUT_WD2@0-
resetting also blocks the watchdog function. Set the watchdog time to m (20...1500)
secondes, with echo of the watchdog time. This command starts the watchdog function and
must be sent within the set watchdog time.

RESET

Switching off the instrument function.

START_X
X=1;2;4

Starting the instrument’s (remote) function

STATUS

Display of status
1S: mode of operation A
2S: mode of operation B
3S: mode of operation C
S0: manual opration without fault
S1: Automatic operation Start (without fault)
S2: Automatic operation Start (without fault)
<0: error code: (-1)

- 1: error 1

- ... (see table)

-31: error 31

-83:wrong parity

-84: unknown instruction

-85:wrong instruction sequence

-86: invalid rated value

-87: not sufficient storage space

STOP_X
X=1;2;4

Switching off the instrument - (remote) function
Variables set with OUT_SP_X are maintained.