IKA LR 1000 basic Package Benutzerhandbuch
Seite 35

35
Abbreviations used:
X,y = numbering parameter (integer number)
m =
variable value, integer
n =
value of variable, floating point number
X = 1
Pt100 thermometer (external temperature sensor)
X = 2 temperature (heating block)
X = 3 safety temperature
X = 4 stirring speed
X = 6 safety stirring speed
Commands
NAMUR Commands
Function
IN_NAME
Input description name
IN_PV_X
X=1;2;3;4;
Reading the real value
IN_SOFTWARE
Input software ID number date, version
IN_SP_X
X=1;2;3;4;6;
Reading the set rated value
IN_TYPE
Input laboratory unit ID
OUT_NAME
Output description name. (Max. 10 characters, default: LR 1000)
OUT_SP_12@n
Setting the WD safety temperature with the echo of the set value
OUT_SP_42@n
Setting the WD safety speed with the echo of the set value
OUT_SP_X n
X=1;2;4;6
Setting the rated value to n
OUT_WD1@m
Watchdog mode 1: When a WD1 event occurs, the heating and shaking functions are
shutdown and message PC 1 is displayed. Set the watchdog time to m (20...1500) seconds,
with echo of the watchdog time. This instruction starts the watchdog function and must be
sent within the set watchdog time.
OUT_WD2@m
Watchdog mode 2: When a WD2 event occurs, the speed setpoint will be set to the WD
safety setpoint speed and the temperature setpoint will be set to the WD safety setpoint
temperature. The PC 2 warning is displayed. The WD2 event can be reset with OUT_WD2@0-
resetting also blocks the watchdog function. Set the watchdog time to m (20...1500)
secondes, with echo of the watchdog time. This command starts the watchdog function and
must be sent within the set watchdog time.
RESET
Switching off the instrument function.
START_X
X=1;2;4
Starting the instrument’s (remote) function
STATUS
Display of status
1S: mode of operation A
2S: mode of operation B
3S: mode of operation C
S0: manual opration without fault
S1: Automatic operation Start (without fault)
S2: Automatic operation Start (without fault)
<0: error code: (-1)
- 1: error 1
- ... (see table)
-31: error 31
-83:wrong parity
-84: unknown instruction
-85:wrong instruction sequence
-86: invalid rated value
-87: not sufficient storage space
STOP_X
X=1;2;4
Switching off the instrument - (remote) function
Variables set with OUT_SP_X are maintained.