Error 12

(Er 12)

External temperature
sensor unplug during
heating control

• External temperature sensor

unplug when heating
function is activated

Stop heating. Exit
error state when
button for heating
control is pressed

• Plug in the external

temperature sensor

Error 14

(Er 14)

Short circuit of
external temperature
sensor or cable

• Short circuit of external

temperature sensor or cable

Heating off

• Check the external

temperature sensor and
cable

Error 17

(Er 17)

The temperature
of safety sensor is
higher than safety
temperature setting

• The temperature of safety

sensor is higher than safety
temperature setting

Heating off

• Increase the safety

temperature setting or allow
cooling down

Error 51

(Er 51)

Mains voltage is too
high

• 220-230V power plug in

100-120V device

Heating off
Motor off

• Switch off the device and

check the mains voltage

Error 52

(Er 52)

Mains voltage is too
low

• 100-120V power plug in

220-230V device

Heating off
Motor off

• Switch off the device and

check the mains voltage

Error 60

(Er 60)

Power is interrupted
when the device is
running

• Power is interrupted when

heating or motor control is
activated in Mode B and C

Exit error state
when any button
is pressed

?

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.

Accessories

T 25 digital disperser

LR 1000.41

shaft receptacle

LR 1000.61

sensor receptacle

LR 1000.65

pH electrode receptacle

Warranty

In accordance with IKA

®

warranty conditions, the warranty

period is 24 months. For claims under the warranty please
contact your local dealer. You may also send the machine
direct to our factory, enclosing the delivery invoice and giving
reasons for the claim. You will be liable for freight costs.

LR 1000.64

pH Electrode

LR 1000.10

anchor Stirrer with PTFE scrapers

LR 1000.20

flow breaker/baffle

The warranty does not cover worn out parts, nor does it
apply to faults resulting from improper use, insufficient care
or maintenance not carried out in accordance with the in-
structions in this operating manual.

for LR 1000 control only. See more accessories on www.ika.com.

Materials in contact with medium/laboratory reactor

Reactor cover

Stainless steel 1.4571

Bottom

Stainless steel 1.4571

Reactor vessel

Borosilicate glass 3.3

Anchor stirrer

Stainless steel 1.4571

Scraper

PTFE/Stainless steel 1.4571

O-ring

FKM

Shaft seal

PTFE