Accessories, Warranty – IKA LR 1000 basic Package Benutzerhandbuch
Seite 39

39
Error 12
(Er 12)
External temperature
sensor unplug during
heating control
• External temperature sensor
unplug when heating
function is activated
Stop heating. Exit
error state when
button for heating
control is pressed
• Plug in the external
temperature sensor
Error 14
(Er 14)
Short circuit of
external temperature
sensor or cable
• Short circuit of external
temperature sensor or cable
Heating off
• Check the external
temperature sensor and
cable
Error 17
(Er 17)
The temperature
of safety sensor is
higher than safety
temperature setting
• The temperature of safety
sensor is higher than safety
temperature setting
Heating off
• Increase the safety
temperature setting or allow
cooling down
Error 51
(Er 51)
Mains voltage is too
high
• 220-230V power plug in
100-120V device
Heating off
Motor off
• Switch off the device and
check the mains voltage
Error 52
(Er 52)
Mains voltage is too
low
• 100-120V power plug in
220-230V device
Heating off
Motor off
• Switch off the device and
check the mains voltage
Error 60
(Er 60)
Power is interrupted
when the device is
running
• Power is interrupted when
heating or motor control is
activated in Mode B and C
Exit error state
when any button
is pressed
?
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
Accessories
T 25 digital disperser
LR 1000.41
shaft receptacle
LR 1000.61
sensor receptacle
LR 1000.65
pH electrode receptacle
Warranty
In accordance with IKA
®
warranty conditions, the warranty
period is 24 months. For claims under the warranty please
contact your local dealer. You may also send the machine
direct to our factory, enclosing the delivery invoice and giving
reasons for the claim. You will be liable for freight costs.
LR 1000.64
pH Electrode
LR 1000.10
anchor Stirrer with PTFE scrapers
LR 1000.20
flow breaker/baffle
The warranty does not cover worn out parts, nor does it
apply to faults resulting from improper use, insufficient care
or maintenance not carried out in accordance with the in-
structions in this operating manual.
for LR 1000 control only. See more accessories on www.ika.com.
Materials in contact with medium/laboratory reactor
Reactor cover
Stainless steel 1.4571
Bottom
Stainless steel 1.4571
Reactor vessel
Borosilicate glass 3.3
Anchor stirrer
Stainless steel 1.4571
Scraper
PTFE/Stainless steel 1.4571
O-ring
FKM
Shaft seal
PTFE