- Wear protective gloves during cleaning the instrument.
- Electrical devices may not be placed in the cleaning agent

for the purpose of cleaning.

- Do not allow moisture to get into the equipment when

cleaning.

- Please consult IKA

®

before using any cleaning or decon-

tamination methods, other than those recommended here.

- The handles can not be place in the cleaning agent for

cleaning.

- The only cleaners or disinfectants that may be used are

those that:

- lie in the pH range 5 - 8,
- contain no corrosive alkalis, peroxides, chlorine compounds,

acids or brine.

Fig. 23

Only clean IKA

®

appliances using these IKA

®

approved

cleaning agents.

Dirt

Cleaning agent

Dyes

Isopropanol

Building materials Water containing detergent, Isopropanol
Cosmetics

Water containing detergent, Isopropanol

Food

Water containing detergent

Fuels

Water containing detergent

Other materials

Please consult IKA

®

Spare parts order

When ordering spare parts, please give:
- machine type
- serial number, see type plate
- item and designation of the spare part see www.ika.com,

spare parts diagram and spare parts list.

Repair

Please send equipment for repair only after it has

been cleaned and is free from any materials which

may constitute a health hazard.
For repairing, please request the “Decontamination

Clearance Certificate” from IKA

®

, or download the print-

out of it from the IKA

®

website www.ika.com.

If you require servicing, return the equipment in its original
packaging. Storage packaging is not sufficient. Please also
use suitable transport packaging.

Error codes

The fault is shown by an error message in the display as following if the error occurs.

Proceed as follows in such cases:

Switch off the device using the main switch

Carry out corrective measures

Restart the instrument

Error code Description

Cause

Effect

Corrective action

Error 3

(Er 3)

Internal temperature of
device is too high

• Ventilation slots or fan housing

is blocked

Heating off
Motor off

• Switch off the unit. Allow it to

cool down and then switch on
again.

• Clean ventilation slots or fan

housing

Error 4

(Er 4)

Motor current is over the
limit

• Motor is blocked or overloaded Motor off

• Reduce the load
• Reduce the setpoint speed

Error 10

(Er 10)

Remote control is
interrupted

• Remote control is interrupt and

relating watchdog function
is on

Heating off
Motor off

• Change watchdong time
• Check relating connection

Error 11

(Er 11)

External temperature
sensor plug during
heating control

• External temperature sensor

plug in when heating function
is activated

Exit error state after
external temperature
sensor is unplug or
button for heating
control is pressed

• Unplug the external temperature

sensor