2 installation, 1 connection dtrans t03 j - type 707030, Installation – JUMO 707030 dTRANS T03 2-Wire/3-Wire Transmitter Operating Manual Benutzerhandbuch

Seite 51

Advertising
background image

7

2 Installation

2.1 Connection dTRANS T03 J - Type 707030/...

Connection for

Terminals

Supply
voltage
7.5 — 30V DC
or
current output
4 — 20mA

+1
-2

R

B

= burden resistance

U

b

= supply voltage

Analog inputs

Resistance
thermometer
in 2-wire
circuit

3
4

R

L

0

as standard

R

L

= lead resistance

per conductor

Setup

R

B

U

b

7.5V

22mA

---------------------

=

Advertising