7 fine calibration, Fine calibration, 3 setup program – JUMO 707030 dTRANS T03 2-Wire/3-Wire Transmitter Operating Manual Benutzerhandbuch
Seite 74
Advertising
30
3 Setup program
3.7.7 Fine calibration
Extras
Fine calibration (or the icon from the toolbar) desig-
nates the adjustment of the output signal of a calibrated transmitter.
The signal can be adjusted within the range of ±0.2mA for current
output and ±0.1V for voltage output.
After calling up the function, the window “Fine calibration” appears:
h
Perform fine calibration by using the arrow buttons and click OK.
You can discard fine calibration with Cancel.
H
When performing fine calibration of the offset and range
end, make sure that the corresponding input signal is pres-
ent.
The fine calibration data will be overwritten at the next data
transfer to the transmitter. In this case, fine calibration has
to be repeated.
Advertising