32

3 Setup program

Preconditions

The following preconditions are necessary for achieving the accura-
cies specified under “Technical data”:

- Precision decade resistor

accuracy: ±0.05%, resolution: 0.01

- Ammeter/voltmeter

accuracy: ±0.05%

־ ±10µA / ±5mV

- Warming-up time: 2min

- Connect transmitter according to the type used

(see Chapter 3.3 “Connection layout for calibration”)