Baumer BA DABU MP4M Benutzerhandbuch

Seite 10

Advertising
background image

DABx_MP4M_BA_V2_3m.doc

19/28

Baumer Electric AG

19.12.07/rut

Frauenfeld, Switzerland

4.3 Electrical connections

Sensor side

PIN Signals Description
1

+ E

+ Bridge excitation

2

+ E

+ Bridge excitation

3

+ E’

Bridge to completion resistor

4

n.c.

5

n.c.

6

+ S

+ Input signal

7

- S

- Output signal

8

n.c.

9

- E’

Bridge to completion resistor

10

- E’

Bridge to completion resistor

11

- E

- Bridge excitation

12

- E

- Bridge excitation

13

Shield

Control side

PIN Signals

Description

1

+ Vs

+ Power supply

2

Vout / Iout

Output signal

3

GND

- Power supply

4

Vout2 / Iout2

Analog output (PEAK)

5

GND

Signal reference

6

Switch S1

Switching output 1

7

Switch S2

Switching output 2

8

Reset

Taring output signal

9

GND

Signal reference

10

Comm_Logic Logic reference

11

Mode_0

Mode 0

12

Mode_1

Mode 1

13

Error

Error switching output

4.4 Bridge connections

2 x ¼ Bridge

Half bridge

Full bridge

12

1

6

7

3

2

10

11

Device
side

Sensor side

E-

E+

S+

S-

E+'

E+

E-'

E-

E-

E+

S+

S-

E+'

E+

E-'

Sensor side

E-

12

1

6

7

3

2

9

11

Device
side

E-

E+

S-

S+

Sensor side

12

1

7

6

Device
side


Voltage supply range: 15 – 33 VDC

Reset-Function: please see chapter 4.7

DABx_MP4M_BA_V2_3m.doc

20/28

Baumer Electric AG

19.12.07/rut

Frauenfeld, Switzerland

4.5 DIP Switches

4.5.1

External DIP switches

These switches can be changed in the running mode.

LCD-Display

Switch

1

2

3

4

ON

Output signal

PEAK

Not occupied

Not occupied

Not occupied

OFF

Output signal

Limit switch*

Not occupied

Not occupied

Not occupied

Factory setting

4.5.2

Internal DIP switches

To have access to the internal DIP switch, the
upper side wall has to be removed before.
Note:
Open device only in supply-free condition.

Proceed as follows:

1. Remove the three marked screws (picture

right hand side).

2. Remove the upper side wall by easy tipping

and lifting.

RESET

Switch

1

2

3

4

ON

Reset active*

Not occupied

Not occupied

Not occupied

OFF

Reset inactive

Not occupied

Not occupied

Not occupied

Factory setting

4.6 Mode function

Mode control

Comment
Galvanically separated and short-circuit
resistant opto couplers serve as switch
elements. They must not be operated
outside the specification (see chapter
5.1, Electrical data)

Example for positive signal change ( Tension )

Mode_0

Mode_1

Reset

Comm_Logic

+Vs

GND

Device side


Mode selection:
The mode selection has to be made
before the supply connecting.
(please see example)

Mode_0 Mode_1 Remark

X

L

Positive signal change ( Tension )

X

H

Negative signal change
( Compression )

Advertising