Sig ofvl peak 8.04v, Sig ufvl peak 8.04v – Baumer BA DABU MP4M Benutzerhandbuch
Seite 13

DABx_MP4M_BA_V2_3m.doc
25/28
Baumer Electric AG
19.12.07/rut
Frauenfeld, Switzerland
4.10 Error case identification
If the measuring value is out of the
measuring range, it is indicated as described:
The error case identification allows the
control of the internal function of the amplifier
and the function of the sensor. Following
Failure can be detected:
•
Interrupted connection between sensor
and amplifier
•
Measuring value out of measurable
range
•
Internal functional error of amplifier
•
No supply voltage on the amplifier
•
Bridge to completion resistor not existing
1. The Error-LED changes from ’green’ to
‘red’
2. The switching output ‘Error’ has been
activated (indicates ‘Low’)
3. Representation on the LCD
SIG OFVL
PEAK 8.04V
SIG UFVL
PEAK 8.04V
If the measuring value is in the measurable
range again, the LED lights green, the
switching output is deactivated and the LCD
shows the correct value.
Note
The error output switch is default set on
‘High’. In error case the switch switches to
‘Low’!
Comment
Should the amplifier remain in error status,
follow following procedure:
Check if the sensor is fully functional
Check if the sensor is connected correctly
Tare the amplifier by activating the Reset
function.
If the amplifier still remains in error status ,
please get in contact with our representative
in your area.
Input signal
Error switching output
Upper range
limit
lower range
limit
High
Low
E
rr
o
r
c
a
s
e
E
rr
o
r
c
a
s
e
P
o
w
e
r
u
p
4.11 Start procedure
After starting the following indications appear:
1. All segments
2. ‘Baumer’
3. All segments
4. ‘DABx’
5. All segments
6. ‘Program Version’
DABx_MP4M_BA_V2_3m.doc
26/28
Baumer Electric AG
19.12.07/rut
Frauenfeld, Switzerland
5 Technical data
5.1 Electrical data
DABU
DABI
General
Voltage supply range
15 – 33 VDC
15 – 33 VDC
Supply current
< 120mA without bridge excitation
< 120mA without bridge excitation
Reset- and mode – Input
(galvanic disconnected)
5 – 33 VDC
<1 VDC
active
inactive
5 – 33 VDC
<1 VDC
active
inactive
Warm-up time
min. 10 min
min. 10 min
Input signal
Bridge excitation
5.1V
±
2%, < 15 mA
5.1V
±
2%, < 15 mA
S/G bridge resistance
≥
350
Ω
≥
350
Ω
Bridge completion resistors
350 Ohm
350 Ohm
Signal input range
According part number key
According part number key
Sample rate
< 1 ms
< 1 ms
Analog output
Output signal
+/- 10 VDC
Load resistance > 10kOhm
4...20mA
apparent ohmic resistance <500Ohm
Characteristic curve deviation
<
±
0.08 % FSR ( <
±
16mV )
<
±
0.15 % FSR ( <
±
24
µ
A )
Resolution Vout / Iout ( 12bit )
< 0.035 % FSR ( 7 mV )
< 0.07 % FSR ( 11
µ
A )
Resolution Vout2 / Iout2 ( 12bit )
< 0.035 % FSR ( 7 mV )
< 0.07 % FSR ( 11
µ
A )
Reset
Zero point reset activ
< ± 10 mV
< ± 10
µ
A
Tare range (reset)
±
6 mV / V
±
6 mV / V
Reset-pulse
> 1 ms
> 1 ms
Reset- settle time
> 5 ms after reset-pulse
> 5 ms after reset-pulse
Reset / operate offset
< ± 10 mV
< ± 15
µ
A
Software
Software version
see chapter 4.11
Display
Display updates
2 per second
2 per second
Resolution
10 mV
10
µ
A
Switching output
switching hysteresis threshold
< 0.25 % FSR ( 50 mV )
< 0.5 % FSR ( 80
µ
A )
Max. load limit switches
50 mA
50 mA
5.2 Mechanical data
Control and sensor connection
13 pin spring block- 0,5 ... 1,5 mm
2
( 20AWG to 16 AWG )
Shield connection
2 pin terminal block
Enclosure
Aluminium / Plastic