Gb 3. specifications – WIKA CTM9100-150 Benutzerhandbuch

Seite 10

Advertising
background image

10

WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

3. Specifications

Approvals and certificates, CTB9100 series

CE conformity
EMC directive

2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity

(industrial application)

Low voltage directive

2006/95/EC, EN 61010-1, safety requirements for electrical equipment for measure-

ment, control and laboratory use

Certificate
Calibration

3.1 calibration certificate per DIN EN 10204

Option: DKD/DAkkS calibration certificate

For further specifications see WIKA data sheet CT 46.30 and the order documentation.

3.3 Multi-function calibrator, model CTM9100-150

Specifications

Model CTM9100-150

Temperature range

-20 ... +150 °C

Accuracy

±0.2 K

±0.3 K

±1 K

±1 K

Operating as a micro calibration bath

Operating as a dry-well temperature calibrator

Operating as an infrared black body source

Operating as a surface temperature calibrator

Stability

±0.05 K

±0.05 K

±0.2 K

±0.2 K

Operating as a micro calibration bath

Operating as a dry-well temperature calibrator

Operating as an infrared black body source

Operating as a surface temperature calibrator

Display resolution

0.1 °C / 0.01 °C

Heating-up time

depending upon usage and area of application

Cool-down time

depending upon usage and area of application

Immersion depth

150 mm

Insert dimensions

Ø 60 x 170 mm

Voltage supply
Power supply

AC 100 ... 240 V, 50/60 Hz

Power consumption

375 VA

Communication
Interface

RS-485

Case
Dimensions (W x H x D)

215 x 425 x 305 mm

Weight

12 kg

Approvals and certificates, model CTM9100-150

CE conformity
EMC directive

2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity

(industrial application)

Low voltage directive

2006/95/EC, EN 61010-1, safety requirements for electrical equipment for measure-

ment, control and laboratory use

Certificate
Calibration

3.1 calibration certificate per DIN EN 10204

Option: DKD/DAkkS calibration certificate

For further specifications see WIKA data sheet CT 41.40 and the order documentation.

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: