13. Heating and cooling times

Infra-red insert

Heating up:

Time

Cooling:

Time

-20 °C to -15 °C

0:30 min

150 °C to 125 °C

2:26 min

-15 °C to -10 °C

0:30 min

125 °C to 100 °C

2:52 min

-10 °C to 0 °C

1:00 min

100 °C to 75 °C

3:36 min

0 °C to 25 °C

2:42 min

75 °C to 50 °C

4:27 min

25 °C to 50 °C

3:06 min

50 °C to 25 °C

6:03 min

50 °C to 75 °C

3:26 min

25 °C to 0 °C

8:59 min

75 °C to 100 °C

4:12 min

0 °C to -10 °C

5:33 min

100 °C to 125 °C

5:38 min

-10 °C to -15 °C

3:31 min

125 °C to 150 °C

9:49 min

-15 °C to -20 °C

4:32 min

13.11 Model CTM9100-150 as surface temperature calibrator
Measuring conditions

All times refer to a Pt100 reference sensor with Ø 3 mm,

L = 300 mm

The reference sensor is inserted to a depth of 111 mm

All times are transition times and do not account for the

required settling time

The measurements were carried out at a room tempera-

ture of approx. 23 °C

It is controlled by an external reference sensor (3 x 300 mm)

Without cap, without covering

13.10 Model CTM9100-150 as an infrared black body source
Measuring conditions

Surface insert

Heating up:

Time

Cooling:

Time

-20 °C to -15 °C

0:46 min

150 °C to 125 °C

3:11 min

-15 °C to -10 °C

0:45 min

125 °C to 100 °C

3:17 min

-10 °C to 0 °C

1:15 min

100 °C to 75 °C

3:51 min

0 °C to 25 °C

2:57 min

75 °C to 50 °C

5:02 min

25 °C to 50 °C

3:16 min

50 °C to 25 °C

6:58 min

50 °C to 75 °C

3:37 min

25 °C to 0 °C

11:55 min

75 °C to 100 °C

4:46 min

0 °C to -10 °C

9:19 min

100 °C to 125 °C

6:18 min

-10 °C to -15 °C

7:44 min

125 °C to 150 °C

10:45 min

-15 °C to -20 °C

12:35 min

All times refer to a Pt100 reference sensor with Ø 3 mm,

L = 150 mm

The reference sensor is inserted to a depth of 51 mm, and

is located under the front surface

All times are transition times and do not account for the

required settling time

The measurements were carried out at a room tempera-

ture of approx. 23 °C

It is controlled by an external reference sensor (3 x 300 mm)

Without cap, without covering