WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

7. Operating elements of the calibrator/micro calibration bath

Cooling gradient SLoF

The cooling gradient, SLoF, is active if the reference temper-

ature is higher than the set temperature.

Only settings that are under the heating power of the calibra-

tor have an effect on the cooling gradient.

Calibrator model

(heating/cooling)

Setting for

SLoF

CTD9100-165

< 5 °C/min

CTB9100-165 / CTM9100-150

with silicone oil 10 CS

< 6 °C/min

CTB9100-165 / CTM9100-150

with distilled water

< 4 °C/min

CTM9100-150 as calibrator

< 4 °C/min

CTM9100-150 as infra-red calibrator < 4 °C/min
CTM9100-150 as surface calibrator

< 4 °C/min

Calibrator model

(heating)

Setting for

SLoF

CTD9100-450 / CTD9100-650

up to 300 °C

300 °C to 100 °C

< 10 °C/min

< 5 °C/min

CTB9100-225

with silicone oil 20 CS

200 °C to 50 °C

50 °C to 30 °C

< 4 °C/min

< 0.5 °C/min

CTB9100-225

with distilled water

90 °C to 50 °C

50 °C to 30 °C

< 2 °C/min

< 0.5 °C/min

The dwell time, dur.t, is active when the set temperature,

SP1, has been reached. Subsequently, the calibrator or micro

calibration bath automatically switches to set temperature

SP2.

CAUTION!

If settings have been made for these three

parameters, the calibrator or the micro

calibration bath will not use the new values

until you either change the set temperature

or until the calibrator or the micro calibration

bath has been turned off and then on again.

Another approach would be to switch the

automatic control off before the parameter

change (see chapter 7.4.2 “Switching off

automatic control”), and then turning it

back on (see chapter 7.4.3 “Switching on

automatic control”).

The heating and cooling gradients and the dwell time can be

set via the parameter level, ’rEG. This is achieved by press-

ing button P for approx. 5 secs., this opens the main menu.

The upper PV display will show OPEr.

In the lower SV display, the LED SET will blink.

If button P is pressed again, this opens the groups level.

The upper PV display will show OPEr.

In the lower SV display, ’SP is shown and additionally LED

SET will blink.

Select the group, rEG, by pressing button 6.

The upper PV display will show OPEr.

In the lower SV display, ’rEGis shown and additionally LED

SET will blink.

’rEGgroup

’SP group

Operator menu OPEr