WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

2. Safety

2.4 Special hazards

WARNING!

For hazardous media such as oxygen,

acetylene, flammable or toxic gases or

liquids, and refrigeration plants, compres-

sors, etc., in addition to all standard regula-

tions, the appropriate existing codes or

regulations must also be followed.

DANGER!

Danger of death caused by electric current

Upon contact with live parts, there is a direct

danger of death.

The instrument may only be installed and

mounted by skilled personnel.

Before replacing the fuse, cleaning,

maintenance/servicing and in the event

of danger, the calibrator or the micro

calibration bath must be disconnected

by unplugging the mains cable from the

power supply outlet.

WARNING!

Residual media in the dismounted instru-

ment can result in a risk to persons, the

environment and the equipment.

Take sufficient precautionary measures.

Thermal fuse

WARNING!

For safety, the calibrator or the micro

calibration bath is fitted with an indepen-

dently-operating temperature fuse, which

cuts out the heater power supply if the

temperature inside the enclosure is too

high. Once the metal block and the liquid

bath have cooled down, send in the

calibrator or the micro calibration bath to

WIKA for examination.

The calibrator and the micro calibration

bath have been designed as measuring

and control instruments. With any opera-

tion of the calibrator/micro calibration

bath not expressly provided for in these

operating instructions, additional protec-

tive measures must be taken.

The calibrator or the micro calibration

bath must NOT be used in explosive

atmospheres (flammable or explosive

atmospheres) without adequate protec-

tion.

WARNING!

Since a malfunction of the calibrator/

micro calibration bath can cause personal

injury or damage to property, the equip-

ment must be protected by additional

electro-mechanical safeguards.

2.5 Safety instructions for using calibration liquids

Calibration liquid, water

Only use distilled water, otherwise the

calibrator tank can become strongly scaled

and soiled.

Calibration liquid, silicone oil

WARNING!

Only use the silicone oil recommended

here.

Before working with silicone oil, read

the material data safety sheet. This can

be requested from the manufacturer or

supplier.

When working with silicone oil, sufficient

ventilation should be ensured in the room

since it can give off pollutants.

Since silicone oil is hygroscopic, after

use, always close the calibration bath

using the transport cover.

The transport cover is fitted with a safety

valve. If the micro calibration bath is

closed when it is still warm, excessive

pressure can build up. In order to avoid

overpressuring, which could lead to

damage to the liquid bath, the safety

valve is activated at a pressure of approx.

2.5 bar. This can lead to hot vapour

escaping.

Wear safety goggles!

Do not allow silicone oil to come into contact

with the eyes.

WARNING!

Risk of burns!

Before transporting or touching the metal

block/liquid bath, make sure that it has

cooled down sufficiently, as there is an acute

fire hazard from both the metal block and

liquid bath and also from the test item.