Gb 3. specifications – WIKA CTM9100-150 Benutzerhandbuch

Seite 9

Advertising
background image

9

WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

3. Specifications

Approvals and certificates, CTD9100 series

CE conformity
EMC directive

2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity

(industrial application)

Low voltage directive

2006/95/EC, EN 61010-1, safety requirements for electrical equipment for measure-

ment, control and laboratory use

Certificate
Calibration

3.1 calibration certificate per DIN EN 10204

Option: DKD/DAkkS calibration certificate

For further specifications see WIKA data sheet CT 41.28 and the order documentation.

3.2 Micro calibration bath, CTB9100 series

Model CTB9100-165

Model CTB9100-225

Temperature range

-35 ... +165 °C

t

amb.

... 225 °C, optional t

amb.

... 255 °C

Accuracy

±0.2 K

±0.3 K

Stability

±0.05 K

±0.05 K

Display resolution

0.1 °C

0.1 °C

Heating time incl. stabilisation

2)

30 min from 20 to 160 °C

20 min from 20 to 225 °C

Cooling time incl. stabilisation

2)

30 min from +20 to -20 °C

30 min from 225 to 50 °C

Immersion depth

150 mm

150 mm

Volume

Approx. 0.6 litres

Approx. 0.6 litres

Tank dimensions

Ø 60 x 165 mm

Ø 60 x 165 mm

Voltage supply
Power supply

AC 100 ... 240 V, 50/60 Hz

AC 230/240 V, 50/60 Hz

(AC 100/115 V, 50/60 Hz)

4)

Power consumption

375 VA

1,000 VA

Power cord

for Europe, AC 230 V

for Europe, AC 230 V

Communication
Interface

RS-485

RS-485

Case
Dimensions

215 x 305 x 425 mm (W x H x D)

150 x 270 x 400 mm (W x H x D)

Weight

12 kg

7.9 kg

2) The reference thermometer with which the measurements are performed, has a diameter of 6 mm.

4) AC 115 V power supply must be specified on the order, otherwise an AC 230 V one will be delivered.

Accessories

Model CTB9100-165

Model CTB9100-225

DC 200.05 silicone oil: -40 ... +130 °C,

FP* = 133 °C

from -35 ... +130 °C, very well usable

not recommended

DC 200.10 silicone oil: -35 ... +160 °C,

FP* = 163 °C

from -35 ... +160 °C, well usable

not recommended

DC 200.20 silicone oil: 10 ... 220 °C,

FP* = 232 °C

not recommended

from 40 ... 225 °C, well usable

DC 200.50 silicone oil: 25 ... 250 °C,

FP* = 280 °C

not recommended

from 80 ... 255 °C, well usable

* FP = Flash point

Advertising